The PC-900PLUS Vital Signs Monitor by Creative Medical is a professional-grade, multi-parameter patient monitoring system built for hospitals, clinics, and medical practices that require reliable, real-time monitoring across a full range of patient types. FDA 510(k) cleared and CE certified, it combines clinical accuracy with an intuitive interface — making it a dependable choice for both acute and general care settings.
Three Configurations to Match Your Clinical Needs
The PC-900PLUS is available in three models, allowing facilities to select only the parameters they need:
- PC-900SN — SpO2, Pulse Rate, and NIBP monitoring
- PC-900SNT — SpO2, Pulse Rate, NIBP, and Temperature monitoring
- PC-900SNTE — SpO2, Pulse Rate, NIBP, Temperature, and ECG monitoring
7-Inch Colour Touchscreen Display
A large 7-inch colour touch screen displays SpO2 and ECG waveforms alongside real-time numeric values simultaneously, giving clinicians an immediate, at-a-glance overview of patient status. The tabletop form factor and user-friendly interface reduce training time and support efficient clinical workflows across busy wards and consulting rooms.
Large-Capacity Data Storage & Trend Review
The PC-900PLUS stores up to 12,000 groups of NIBP measurements, 3,000 groups of SpO2 readings, and event records — all reviewable in waveform, list, or trend graph format directly on the device. This depth of historical data supports clinical decision-making and patient progress tracking over time. Data export and software upgrades are handled via USB cable, keeping administration simple.
Compatible with Industry-Leading Accessories
The PC-900PLUS supports Nellcor-compatible SpO2 sensors, SunTech NIBP technology, and infrared ear temperature probes. Optional accessories include a printer, trolley, and clamp — enabling flexible deployment across wards, theatres, outpatient settings, and emergency departments.
Technical Specifications
NIBP (Non-Invasive Blood Pressure)
| Parameter | Detail |
|---|---|
| Technique | Oscillometric |
| Measurement Accuracy | Max mean difference: ±5mmHg; Max SD: 8mmHg |
| Systolic Pressure | Adult: 40–275mmHg / Pediatric: 40–200mmHg / Neonate: 40–135mmHg |
| Diastolic Pressure | Adult: 10–210mmHg / Pediatric: 10–150mmHg / Neonate: 10–95mmHg |
| Mean Arterial Pressure | Adult: 20–230mmHg / Pediatric: 20–165mmHg / Neonate: 20–110mmHg |
| Measurement Mode | Manual, Auto, STAT |
| Overpressure Protection | Adult: 300mmHg / Pediatric: 240mmHg / Neonate: 150mmHg |
SpO2
| Parameter | Detail |
|---|---|
| Transducer | Dual-wavelength LED |
| Measuring Range | 35%–100% |
| Measuring Accuracy | ±3% (70%–100% range) |
| Low Perfusion Performance | As low as 0.4% |
Pulse Rate
| Parameter | Detail |
|---|---|
| Measuring Range | 30bpm–240bpm |
| Measuring Accuracy | ±2bpm or ±2%, whichever is greater |
ECG
| Parameter | Detail |
|---|---|
| Input Dynamic Range | ±0.5mVp–±5mVp |
| HR Measuring Range | 15bpm–350bpm |
| HR Measuring Accuracy | ±1% or ±2bpm, whichever is greater |
| HR Alarm Delay Time | ≤10s |
| Sensitivity Selection | ×1/4, ×1/2, ×1, ×2 |
| Sweeping Speed | 25mm/s |
| ECG Noise Level | ≤30µVp-p |
| ECG Input Loop Current | ≤0.1µA |
| Differential Input Impedance | ≥10MΩ |
| Common-Mode Rejection Ratio | ≥105dB |
Temperature
| Parameter | Detail |
|---|---|
| Measuring Range | 21.0°C–50.0°C |
| Measuring Accuracy | ±0.2°C (25°C–45°C range) |
General
| Parameter | Detail |
|---|---|
| Display | 7-inch colour touch screen |
| Power Supply | AC: 100–240V, 50/60Hz |
| Built-in Battery | 11.1V / 4400mAh Li-ion |
| Certifications | FDA 510(k), CE |
| Available Accessories | ECG cable, NIBP cuff, Skin temp probe, Infrared ear probe, SpO2 sensor |
| Optional Accessories | Nellcor SpO2, Infrared ear temp probe, Printer, Clamp, Trolley |